For intelligent wearable specially designed testing equipment, integrate professional testing software and test instruments. The entire test process can be achieved throughout the process.
Equipment features
• Powerful and complete monitoring software and high precision tester (CMRR3.0, AECG100);
• IPC storage complete detection parameters and output data;
• It can be matched according to the needs to become a fully automated testing equipment.
Equipment parameters
category | project | Specification |
equipment | Voltage | 220V16A |
air pressure | 0.4~0.7Mpa | |
Industrial computer | iPC+WIN7 | |
physical dimension | 325*690*430mm | |
weight | 45kg | |
CMRR3.0+ common modeSuppression Ratio Tester | Supply voltage | 0.5/2.828/20/2(70.71)(Vrms)±1% |
Common mode point voltage | 1.414/10/1/(35.355)(Vrms)±1%(0.25Vrms±2%) | |
frequency | 50/60/100/120(Hz)±1% | |
AECG100 Pulse BloodOxygen ECG Tester | Main output voltage resolution | 2.5uv (DAC resolution) |
Main output voltage accuracy | ±1% for amplitudes of 0.5mVpp or higher | |
Frequency/Pulse Repetition Rate Accuracy | ±1% | |
Pulse amplitude range | ±1mV~1000mV | |
±2mV pulse; ±1% | ||
Pulse Amplitude Accuracy | >2mV pulse, <-2mV; ±10% | |
Pulse width range | 0.1ms~2ms | |
Pulse width resolution | 0.1 ms | |
Pulse width accuracy | ±5 microseconds |